DAR File No. 39016

This rule was published in the January 15, 2015, issue (Vol. 2015, No. 2) of the Utah State Bulletin.


Environmental Quality, Radiation Control

Section R313-28-31

General and Administrative Requirements

Notice of Proposed Rule

(Amendment)

DAR File No.: 39016
Filed: 12/19/2014 09:54:30 AM

RULE ANALYSIS

Purpose of the rule or reason for the change:

The reason for the change is based on a petition for rulemaking. The petitioner requested that hand-held medical x-ray systems be allowed for use on patients. During a review of the petition, the Utah Radiation Control Board (Board) determined that rulemaking proceedings should be initiated to address the petition.

Summary of the rule or change:

The proposed change to Section R313-28-31 adds a new Subsection R313-28-31(6). This subsection outlines the requirements that apply to the use of hand-held medical x-ray equipment, as well as some regulatory exclusions. The requirements specify the actions needed to protect the operator of the x-ray system and members of the public. The rule also requires that each operator of hand-held x-ray equipment complete training supplied by the manufacturer and keep records of the training.

State statutory or constitutional authorization for this rule:

  • Section 19-3-108
  • Subsection 19-3-104(4)

Anticipated cost or savings to:

the state budget:

There is no additional cost to the state budget because a registration and radiation safety inspection program exists for all types of x-ray systems. Allowing the use of hand-held medical x-ray systems will generate some additional revenue for the registration and inspection of such systems. Because the number of systems to be purchased by practitioners is unknown, the amount of this additional revenue is also unknown.

local governments:

Local government health clinics may have a need to acquire a hand-held medical x-ray system. The cost for registering the system is $35 per year and the inspection cost is $105 per inspection.

small businesses:

The Division of Radiation Control is currently regulating many healing arts facilities that are a small business. If the facility acquires a hand-held medical x-ray system, the cost for registering the system is $35 per year and the inspection cost is $105 per inspection.

persons other than small businesses, businesses, or local governmental entities:

The Division of Radiation Control is currently regulating many healing arts facilities that fit within the definition of person. The cost to all persons that acquire a hand-held medical x-ray system is $35 per year for registering the system and $105 per radiation safety inspection.

Compliance costs for affected persons:

There is compliance cost for a protective apron. The average cost for a protective apron is $150. If it is assumed that a facility needs to protect the x-ray system operator and one member of the public during each diagnostic procedure, then the facility must obtain two protective aprons at a total cost of $300.

Comments by the department head on the fiscal impact the rule may have on businesses:

There exists a regulatory program within the department to register x-ray machines and to perform a radiation safety inspection of x-ray systems. The annual registration fee is $35 per x-ray tube and the inspection fee is $105 per inspection. Facilities that own a hand-held medical x-ray system will be required to pay these fees. A regulated facility with a hand-held medical x-ray system will have to obtain protective aprons to protect the operator of the x-ray system and members of the public who may be exposed to the radiation. The cost for a protective apron is minimal and is about $150 per apron.

Amanda Smith, Executive Director

The full text of this rule may be inspected, during regular business hours, at the Division of Administrative Rules, or at:

Environmental Quality
Radiation ControlRoom Third Floor
195 N 1950 W
SALT LAKE CITY, UT 84116-3085

Direct questions regarding this rule to:

  • Craig Jones at the above address, by phone at 801-536-4264, by FAX at 801-533-4097, or by Internet E-mail at [email protected]
  • John Hultquist at the above address, by phone at 801-536-4263, by FAX at 801-536-4250, or by Internet E-mail at [email protected]

Interested persons may present their views on this rule by submitting written comments to the address above no later than 5:00 p.m. on:

02/17/2015

This rule may become effective on:

03/17/2015

Authorized by:

Rusty Lundberg, Director

RULE TEXT

R313. Environmental Quality, Radiation Control.

R313-28. Use of X-Rays in the Healing Arts.

R313-28-31. General and Administrative Requirements.

(1) Persons shall not make, sell, lease, transfer, lend, or install x-ray equipment or the accessories used in connection with x-ray equipment unless the accessories and equipment, when properly placed in operation and properly used, will meet the applicable requirements of these rules.

(2) The registrant shall be responsible for directing the operation of the x-ray machines which are under the registrant's administrative control. The registrant or registrant's agent shall assure that the requirements of R313-28-31(2)(a) through R313-28-31(2)(i) are met in the operation of the x-ray machines.

(a) An x-ray machine which does not meet the provisions of these rules shall not be operated for diagnostic purposes, when directed by the Director.

(b) Individuals who will be operating the x-ray equipment shall be instructed in the registrant's written radiation safety program and be qualified in the safe use of the equipment. Required operator qualifications are listed in R313-28-350.

(c) The registrant of a facility shall create and make available to x-ray operators written safety procedures, including patient holding and restrictions of the operating technique required for the safe operation of the x-ray systems. Individuals who operate x-ray systems shall be responsible for complying with these rules.

(d) Except for individuals who cannot be moved out of the room and the patient being examined, only the staff and ancillary personnel or other individuals needed for the medical procedure or training shall be present in the room during the radiographic exposure and shall be positioned as follows:

(i) individuals other than the patient shall be positioned so that no part of the body will be struck by the useful beam unless protected by not less than 0.5 mm lead equivalent material;

(ii) the x-ray operator, other staff, ancillary personnel and other individuals needed for the medical procedure shall be protected from primary beam scatter by protective aprons or barriers unless it can be shown that by virtue of distances employed, EXPOSURE levels are reduced to the limits specified in R313-15-201; and

(iii) patients who are not being examined and cannot be removed from the room shall be protected from the primary beam scatter by whole body protective barriers of not less than 0.25 mm lead equivalent material or shall be so positioned that the nearest portion of the body is at least two meters from both the tube head and nearest edge of the image receptor.

(e) For patients who have not passed reproductive age, gonad shielding of not less than 0.5 mm lead equivalent material shall be used during radiographic procedures in which the gonads are in the useful beam, except for cases in which this would interfere with the diagnostic procedure.

(f) Individuals shall be exposed to the useful beam for healing arts purposes only when the exposure has been specifically ordered and authorized by a licensed practitioner of the healing arts after a medical consultation. Deliberate exposures for the following purposes are prohibited:

(i) exposure of an individual for training, demonstration or other non-healing arts purposes; and

(ii) exposure of an individual for the purpose of healing arts screening except as authorized by R313-28-31(2)(i).

(g) When a patient or film must be provided with auxiliary support during a radiation exposure:

(i) mechanical holding devices shall be used when the technique permits. The written procedures, required by R313-28-31(2)(c), shall list individual projections where mechanical holding devices can be utilized;

(ii) written safety procedures, as required by R313-28-31(2)(c), shall indicate the requirements for selecting an individual to hold patients or films and the procedure that individual shall follow;

(iii) the individual holding patients or films during radiographic examinations shall be instructed in personal radiation safety and protected as required by R313-28-31(2)(d)(i);

(iv) Individuals shall not be used routinely to hold film or patients;

(v) In those cases where the patient must hold the film, except during intraoral examinations, portions of the body other than the area of clinical interest struck by the useful beam shall be protected by not less than 0.5 mm lead equivalent material; and

(vi) Facilities shall have protective aprons and gloves available in sufficient numbers to provide protection to personnel who are involved with x-ray operations and who are otherwise not shielded.

(h) Personnel monitoring. Individuals who are associated with the operation of an x-ray system are subject to the applicable requirements of R313-15.

(i) Healing arts screening. Persons proposing to conduct a healing arts screening program shall not initiate the program without prior approval of the Director. When requesting approval, that person shall submit the information outlined in R313-28-400. If information submitted becomes invalid or outdated, the Director shall be notified immediately.

(3) Maintenance of records and information. The registrant shall maintain at least the following information for each x-ray machine:

(a) model numbers of major components;

(b) record of surveys or calculations to demonstrate compliance with R313-15-302, calibration, maintenance and modifications performed on the x-ray machine; and

(c) a shielding design report for the x-ray suite which states assumed values for workload and use factors and includes a drawing of surrounding areas showing assumed values for occupancy factors.

(4) X-ray records. Facilities shall maintain an x-ray record containing the patient's name, the types of examinations, and the dates the examinations were performed. When the patient or film must be provided with human auxiliary support, the name of the human holder shall be recorded. The registrant shall retain these records for three years after the record is made.

(5) Portable or mobile equipment shall be used only for examinations where it is impractical to transfer the patient to a stationary radiographic installation.

(6) Hand-held medical x-ray systems. X-ray equipment designed to be hand-held shall comply with Section R313-28-31, excluding Subsection R313-28-31(5), and R313-28-52, excluding Subsections R313-28-52(8)(b)(i) and (ii).

(a) When operating hand-held equipment for which it is not possible for the operator to remain at least six feet from the x-ray machine during x-ray exposure, protective aprons of at least 0.5 millimeter lead equivalence shall be provided for the operator to protect the operator's torso and gonads from backscatter radiation;

(b) In addition to the dose limits in R313-15-301, operators of hand-held x-ray equipment shall ensure that members of the public that may be exposed to scatter radiation or primary beam transmission from the hand-held device are not exposed above 2 milliroentgen per hour;

(i) Operators will ensure that members of the public likely to be exposed to greater than 2 milliroentgen per hour will be provided protective aprons of at least 0.5 millimeter lead equivalence or are moved to a distance such that the exposure rate to the individual is below 2 milliroentgen per hour; and

(c) In addition to the requirements of Subsection R313-28-350(1), each operator of hand-held x-ray equipment shall complete the training program supplied by the manufacturer prior to using the x-ray unit. Records of training shall be maintained on file for examination by an authorized representative of the Director.

([6]7) Procedures and auxiliary equipment designed to minimize patient and personnel exposure commensurate with the needed diagnostic information shall be utilized.

(a) The speed of the screen and film combinations used shall be the fastest speed consistent with the diagnostic objective of the examinations. Film cassettes without intensifying screens shall not be used for routine diagnostic radiological imaging, with the exception of standard film packets for intra-oral use in dental radiography. If the requirements of R313-28-31(6)(a) cannot be met, an exemption may be requested pursuant to R313-12-55.

(b) The radiation exposure to the patient shall be the minimum exposure required to produce images of good diagnostic quality.

(c) X-ray systems, other than fluoroscopic, computed tomography, dental or veterinary units, shall not be utilized in procedures where the source to patient distance is less than 30 centimeters.

 

KEY: dental, X-rays, mammography, beam limitation

Date of Enactment or Last Substantive Amendment: [October 15, 2013]2015

Notice of Continuation: September 23, 2011

Authorizing, and Implemented or Interpreted Law: 19-3-104; 19-3-108

 


Additional Information

More information about a Notice of Proposed Rule is available online.

The Portable Document Format (PDF) version of the Bulletin is the official version. The PDF version of this issue is available at https://rules.utah.gov/publicat/bull-pdf/2015/b20150115.pdf. The HTML edition of the Bulletin is a convenience copy. Any discrepancy between the PDF version and HTML version is resolved in favor of the PDF version.

Text to be deleted is struck through and surrounded by brackets ([example]). Text to be added is underlined (example).  Older browsers may not depict some or any of these attributes on the screen or when the document is printed.

For questions regarding the content or application of this rule, please contact Craig Jones at the above address, by phone at 801-536-4264, by FAX at 801-533-4097, or by Internet E-mail at [email protected]; John Hultquist at the above address, by phone at 801-536-4263, by FAX at 801-536-4250, or by Internet E-mail at [email protected].  For questions about the rulemaking process, please contact the Division of Administrative Rules.